Reader Unit Set for SENSOR RESERVOIR
Message displayed
Wrong SD card inserted!
Remove SD card!
:
Measurement without data sto-
rage possible
- or -
Insert SD card with correct ID
Wrong SD card inserted!
Remove SD card!
:
Measurement without data sto-
rage possible
- or -
Insert SD card with correct ID
XXXXXXXXXX!
Wrong implant - restart
measurement!
Switching to another implant
during continuous measure-
ment not possible!
Unit-temperature incorrect
:
Unit temperature from 10°C to
40°C valid!
Internal voltage incompatible
SD-card faulty!
:
Measurement w/o storage of
data possible
SD-card faulty!
SD-card inserted!
Restart measurement.
:
Storage of reading after restart
of measurement possible!
SD-card missing!
:
Insert SD card!
Cause
Measurement starts without SD
card. During measurement, an
SD card not designated for use
with that implant is used during
measurement
SD card with different ID to the
implant has been used
After the start of a continuous
measurement, data from another
implant have been received
Temperature in the reader unit
exceeds calibrated interval
Internal voltage of device too
high/ too low
SD card not writeable or
readable (dirt, corrosion, faulty
contacts)
SD card not readable (dirt, corro-
sion, faulty contacts)
Measurement starts without SD
card. During measurement, the
designated SD card for use with
that implant is inserted
No SD card inserted in data ma-
nagement mode
INSTRUCTIONS FOR USE |
Fault detection/rectification
Insert the SD card designated for use with
the SENSOR RESERVOIR / SENSOR
PRECHAMBER into the reader unit.
The implant (see patient card) is associated
with the respective SD card via the identifica-
tion number (ID).
Insert the SD card designated for use with
the SENSOR RESERVOIR / SENSOR
PRECHAMBER into the reader unit.
The implant (see patient card is associated
with the respective SD card via the identifica-
tion number (ID).
Increase the distance between the two im-
plants.
The reader unit can only be used at device
temperatures from 10°C to 40°C.
Current measurement is stopped.
Reader unit switches off automatically after
20 seconds.
Contact Technical Service.
Use of implant's internal calibration data.
Data are not saved.
Check SD card for damage or dirt.
Start measurement again.
Insert SD card.
GB
51