| INSTRUCTIONS FOR USE
GB
Message displayed
SD-card missing!
:
Measurement w/o storage of
data possible
- or -
Insert SD-card with
correct ID XXXXXXXXXX!
SD-card not readable!
:
Measurement w/o data storage
possible!
SD-card not readable!
:
Measurement w/o storage of
data possible!
SD-card memory full.
Option to measure w/o saving
SD-card memory full
SD-card memory almost full
System error
:
Ambient pressure not readable
System error
- Incompatibility
System error
- Antenna incompatible
System error
- ID data incorrect
System error
- Implant voltage out of range
52
Cause
SD card not inserted
Insertion of wrongly formatted or
unformatted SD card
SD card missing
- or -
SD card ID not readable
- or -
SD card contains no calibration
data
SD card memory full (100%)
during continuous measurement
SD card memory full (100%)
during continuous measurement
SD card memory almost full
(99%) during continuous or fast
measurement
Barometric (BARD) pressure not
readable
Device hardware and software
version not compatible
Hardware version of antenna and
reader unit not compatible
Implant identification data
corrupted
Implant voltage outside of
permissible range
Reader Unit Set for SENSOR RESERVOIR
Detection/remedy
Insert SD card designated for use with the
SENSOR RESERVOIR / SENSOR
PRECHAMBER into the reader unit.
The implant (see patient card) is associated
with the respective SD card via the identifica-
tion number (ID).
Insert SD card designated for use with the
SENSOR RESERVOIR / SENSOR
PRECHAMBER into the reader unit.
The implant (see patient card) is associated
with the respective SD card via the identifica-
tion number (ID).
Insert SD card designated for use with the
SENSOR RESERVOIR / SENSOR
PRECHAMBER into the reader unit.
Measurement with data storage possible, but
with limited accuracy.
Delete any data no longer required, measure-
ment without data storage possible.
Delete any data no longer required.
Delete any data no longer required.
Device function locked. Deactivate and reac-
tivate device.
If the error persists, contact Technical Service.
Device function locked. Deactivate and re-
activate device. If the error persists, contact
Technical Service.
Device function locked. Deactivate and re-
activate device. If the error persists, replace
antenna and/or contact Technical Service.
Measurement is stopped. Deactivate and re-
activate device. If the error persists, contact
Technical Service.
Measurement is stopped. Deactivate and re-
activate device. If the error persists, contact
Technical Service.