| INSTRUCTIONS FOR USE
GB
TECHNICAL DATA
Description
Voltage range
Reader unit
Power supply unit
Power consumption
Reader unit
Power supply unit
Operating frequency
Pressure measuring range
Temperature range for pressure measurement
Functional distance between antenna and SENSOR
RESERVOIR / SENSOR PRECHAMBER
Battery life
Battery charging cycle
Battery self-discharge
Flammability category of housing
Moisture protection or protection against ingress of liquids
Reader unit
Antenna
Power supply unit
Durability - impact test
Durability - drop test
Weight:
Reader unit
Antenna
Power supply unit
Dimensions (W x H x D)
Reader unit
Antenna (without cable)
Power supply unit
Protection class (power supply unit)
Conformity to standards
EMC
54
Reader Unit Set for SENSOR RESERVOIR
Specifications and standards
6 V (DC)
100-240 V (50–60 Hz)
1.4 A (DC)
0.25 A (50–60 Hz)
133 kHz
-66.67 hPa bis 133.33 hPa±5.32 hPa
10°C to 40°C (Reader unit)
20°C to 39°C (SENSOR RESERVOIR /
SENSOR PRECHAMBER)
10-30 mm
5 years minimum
at least 250 charging cycles
Remaining charge after 3 months (storage) > 70%
UL 94 HB
IP44
IP44
IP40
in accordance with IEC 60601-1 (3rd edition) 15.3.2
in accordance with IEC 60601-1 (3rd edition) 15.3.4.1
0.600 kg
0.215 kg
0.115 kg
144 x 270 x 65 mm
100 x 250 x 25 mm
55 x 65 x 70 mm
II
IEC/DIN EN60601-1/ EN 45502-1
EN 60601-1-2:2007