No Response
Swipe the screen to the right to access the parameters:
-
Stimulus type: Tone or Warble. Default: Warble.
-
Stimulus mode: Continuous or Pulsed 1 Hz. Default: Continuous.
-
Default intensity: Sets the default intensity of the stimulus. Default:
40 dB HL.
-
Maintain intensity: Maintain the level after changing the frequency.
Default: disabled.
-
Interrupter mode: Allows the interrupter key to be used as a button
(stimulation is active when the key is pressed) or switch (the first key
pressure activates the stimulus, the second one deactivates it). Default:
button.
-
Automatic frequency jump: Enables/disables automatic frequency jump
after a value is stored. Default: disabled.
-
Frequency selection: Access the frequency selection screen to
individually enable/disable the frequencies to be tested. Default value:
all frequencies enabled.
"Bone conduction" license
3.3.1
The following controls and information are available only when the "Bone
conduction" license is enabled:
Icon
Select the transducer
Enable tracking (keep the same difference in dB between
stimulus and masking)
Frequency not tested
Operation
EN