The use of accessories, transducers and cables other than those
specified, except for transducers and cables sold by the
manufacturer as spare parts for internal components, may result in
increased emissions or decreased electromagnetic immunity of the
device and result in improper operation.
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the device, including cables
specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Anyone connecting additional equipment is responsible for making sure the
system complies with the IEC 60601-1-2 standard.
The device has no ESSENTIAL PERFORMANCE as related to IEC 60601-1.
Note: All instructions necessary for maintaining compliance with regard to
electromagnetic compatibility can be found in the maintenance section in this
manual. No further steps are required.
Guidance and manufacturer's declaration – electromagnetic emissions
Triangle is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that they are used in such an environment.
Emissions test
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations /
flickers emissions
IEC 61000-3-3
36
Compliance
Triangle uses RF energy only for its internal
function. Therefore, its RF emissions are very
Group 1
low and not likely to cause any interference in
nearby electronic equipment.
Class B
Triangle is suitable for use in professional
Class A
healthcare facility environments and directly
connected to the public low-voltage power supply
network.
Complies
Electromagnetic environment guidance