Descargar Imprimir esta página

Technical Specifications - Inventis TRIANGLE Manual

Audiometer

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 186
Performance
Electrical safety
EMC
Duration of calibration
Battery
Duration
Auto-off time
Stand-by time
Recharge time
Max power consumption
External power adapter
Operation
Transport and storage
Warm-up time

Technical Specifications

APPLICABLE STANDARDS
IEC 60645-1 type 4, ANSI S3.6 type 4
IEC 60601-1, Class II, Type B
IEC 60601-1-2
CALIBRATION
12 months
POWER SUPPLY
Rechargeable Li-Ion, standard 18650, 3.7V 2.6Ah
Minimum 12h (continuous use)
5 minutes
1 minute
From PC, standard USB port: 10h max
From dedicated power adapter: 3h max
7W
Mod. FOX6-XM-USB, USB type, IEC 60601-1
compliant. Input 100-240Vac 50/60Hz, 0.3-0.15A,
Output 5Vdc 1.4A
AMBIENT CONDITIONS
Temperature: 15°C to 35°C
Relative humidity: 30% to 90% (no condensation)
Pressure: 700 mbar to 1060 mbar
Temperature: -10°C to 50°C
Relative humidity: 0% to 90% (no condensation)
Pressure: 500 mbar to 1060 mbar
1 minute
Appendix A
27
EN

Publicidad

loading