Instructions for Use
Before use, please read the following instructions. Should you have any queries, please contact
Frontier Therapeutics Limited on +44 (0) 330 460 6030 or visit our website
www.frontier-group.co.uk.
Platform Installation and Operating Guide
To ensure the system is correctly installed and functions smoothly, follow this step-by-step guide.
1.
Remove all items from packaging and inspect for damage.
2.
Remove the mattress and place the rolled platform at the foot end of the bed, roll out until it is
fully extended and positioned centrally to the bed frame.
3.
Ensure the air cells within the platform are secured in position via the popper fastenings by
unzipping the cover and visually inspecting.
4.
If required, re-fit the cover and close the zip ensuring the air cell hoses are fed through the
opening at the foot end of the cover.
Please Note: The platform cover should be free to move, allowing each side of the platform
unrestricted room to inflate.
5.
Using the quick connectors, connect the air inflation hoses to the air cell hoses - 'Red to Red' -
ensuring the hoses are not twisted or kinked.
6.
Check the Cardiopulmonary Resuscitation deflation valves are securely in place to ensure no
air leaks.
7.
Replace the mattress on top of the platform. DO NOT strap or fasten the mattress to the bed
frame.
8.
Using the bed hooks, hang the control unit on the bed footboard.
9.
Using the quick connectors, connect the air inflation hoses to the control unit - 'Black to Control
Unit' - ensuring the hoses are not twisted or kinked.
10.
Connect the control unit to mains power via the lead supplied.
A. The platform should be
placed directly onto the bed
frame and underneath the
existing mattress.
D. Check the Cardiopulmonary
Resuscitation deflation valves
are securely in place to ensure
no air leaks.
B. Inspect the platform for
visible signs of damage.
E. Using the bed hooks, hang
the control unit on the bed
footboard.
11
C. Using the quick connectors,
connect the air inflation hoses
to the air cell hoses - 'Red to
Red' - ensuring the hoses are
not twisted or kinked.
F. Using the quick connectors,
connect the air inflation hoses
to the control unit - 'Black
to Control Unit' - ensuring
the hoses are not twisted or
kinked.