System Error Trouble Shooting Guide
Problem
No LEDs lit;
•
no audible
alarm heard.
•
Air-leak
•
alarm
LED (6) lit
permanently
yellow with
an audible
alarm.
Blockage
•
alarm
LED (7) lit
permanently
yellow with
an audible
alarm.
Service
•
LED (8)
permanently
lit amber.
Interface
•
panel is lit
but unre-
sponsive.
An audible
•
alarm is
sounding; no
•
LEDs lit.
•
Do not open the control unit. Opening the unit could cause personal injury or equipment damage.
NB. The Toto Touch control unit self calibrates on start up.
Cause
Control unit may not
be attached to a power
source.
The fuse may need
replacing.
Pressure has fallen
below the minimum
operational require-
ments.
Airflow obstruction
between the control
unit and the platform.
Service required.
Interface panel is
locked.
Loose or absent power
lead.
Mains power acciden-
tally turned off.
Mains power failure.
Initial Checks / Trouble Shooting
1.
Check that the mains power outlet is live and
active.
2.
Check the control unit is switched on.
3.
If following points 1 and 2 does not resolve
the problem, switch off and unplug the
control unit:
- Check the mains plug fuse (3A).
- Check the control unit fuse (1A).
4.
Restart the control unit.
If the problem persists, contact Frontier
Therapeutics Limited Tel: +44 (0) 330 460 6030
for assistance.
1.
Press the alarm pause button.
2.
Check the Cardiopulmonary Resuscitation
valve is closed and correct.
3.
Check all air cells and hoses for any air
leakage.
4.
Resolve and wait for the alarm to reset.
If the problem persists, contact Frontier
Therapeutics Limited Tel: +44 (0) 330 460 6030
for assistance.
1.
Press the alarm pause button.
2.
Check for twists or kinks in the air hoses
between the platform and control unit.
3.
Resolve and wait for the alarm to reset.
If the problem persists, contact Frontier
Therapeutics Limited Tel: +44 (0) 330 460 6030
for assistance.
Please contact Frontier Therapeutics Customer
Services for support Tel: +44 (0) 330 460 6030
The Control Unit continues to function even
though the service LED remains lit.
1.
Check that screen lock LED is lit amber (9).
2.
Press and hold the alarm mute button for
four seconds to deactivate.
If the problem persists, contact Frontier
Therapeutics Limited Tel: +44 (0) 330 460
6030 for assistance.
1.
Reconnect mains power lead to the control
unit.
2.
Check the mains power wall switch is turned
on.
3.
Attempt to restart the control unit.
If the problem persists, contact Frontier
Therapeutics Limited Tel. +44 (0) 330 460
6030 for assistance.
18