Programmes; Programme Settings - TensCare Sure PRO Instrucciones De Uso

Ocultar thumbs Ver también para Sure PRO:

Publicidad

Idiomas disponibles

Idiomas disponibles

• Pulse Width (measured in μs -
millionths of a second)
The Sure PRO unit has pulse widths of
50 to 500 μs. Generally speaking, the
higher the pulse width, the more
"aggressive" the stimulation feels,

10. PROGRAMMES

10.1. PROGRAMME SETTINGS

Programme
Frequency
(Hz)
1
STRE 1
50
2
STRE 2
35
3
URGE
10
4
MIXED
10/50
5
URGE 2
10
6
SENS
3/10/20/30/40
7
PFW A
See the table below
8
PFW B
See the table below
9
ENDUR
20
10 PAIN
4
11 TIBN 1
10
12 TIBN 2
20
13 CSTM 1
4-90
14 CSTM 2
4-90
15 CSTM 3
4-90
PFW
Frequency
Pulse Width (µs)
(Hz)
A
20
250 µs fixed
20
250 µs stepped
3
250 µs fixed
10
500 µs fixed
20
250 µs stepped
35
250 µs stepped
45
250 µs stepped
10
500 µs fixed
B
20
250 µs fixed
20
250 µs stepped
3
250 µs fixed
if the pulse width is set high enough, it
will usually elicit a muscle contraction,
which is required for an effective toning
of the pelvic floor muscles.
Pulse Width
Ramp
(µs)
(sec)
300
1
250
2
200
1
200/300
200
Constant
250/200
250
2
200
Constant
200
Constant
200
Constant
50-400
1-10
50-400
1-10
50-400
1-10
Work(sec)
4
4
4
4
4
4
4
4
6
6
6
16
Work
Rest
Default
(sec)
(sec)
duration
(min)
5
10
20
3
6
20
5
10
20
30
Continuous
25
5
10
20
20
20
30
1-20
0-30
5-60
1-20
0-30
5-60
1-20
0-30
5-60
Rest (sec)
Default
duration (min)
6
2
6
5
4
5
4
15
6
10
8
5
8
5
4
5
Total
52 min
8
2
8
5
6
5