Ramp (rise)
This is the time in seconds taken to
move up and down between zero and
the set stimulation strength. The Sure
PRO can be adjusted from 1 to 10
seconds in steps of 0.5 sec.
Work
This is the time in seconds that muscle
is stimulated (not including Ramp time).
The Sure PRO offers a range of work
periods from 1 to 20 sec.
Rest
This is the time in seconds at zero
strength in between stimulation. The
Sure PRO offers a range of rest periods
from 0 to 30 sec. The EMS programmes
use an Active Rest - low frequency
pulses help to clear metabolites in
between Work periods.
Note: If Rest is set to 0, Ramp
is also zero and output is
continuous.)
Work/Rest Ratio
Warning: The warning triangle
is displayed if the Rest
period is less than: Rest
Time=Work
16.66)/16.66 seconds.
This is because muscle fibres can only
activate a limited number of times a
time*(WorkHz-
23
minute (about 1000) without becoming
fatigued.
If the warning triangle is displayed at
any other time, see the troubleshooting
in section 20.

11. CONTENT

The pack contains:
• 1 x Sure PRO, continence stimulator
unit
• 2 x Lead wires (L-ST2)
• 1 x Liberty vaginal probe (X-VP)
• 4 x 50x50mm electrode pads (E-
CM5050)
• 1 x Charging adaptor (EU/UK)
• 1 x Detachable belt clip
• 1 x Storage pouch
• 1 x Instructions for use