Included in the package
Description
PULSE OXIMETER (instrument)
PULSE OXIMETER Operator's Manual
lanyard
Temperature probe: Disposable Skin-surface
Temperature Probe, 0.9m.
Installation and Verification
● Battery installation
Caution: The Pulse Oximeter does not operate with dead batteries and can not be
powered by external power source does not input outer power.
Install new batteries.
1. Unplug all accessories from the Pulse Oximeter, and press the menu bar to
access the Setting Interface, turn the PULSE OXIMETER off. See table 1.
2. Remove the battery rear cover towards the direction indicated on the cover.
3. Insert two "AAA" size batteries, making sure the battery's positive and
negative poles are correctly oriented in the battery compartment as shown in
Figure 4.
4. Close the battery rear cover.
● Performance Verification
1. Performance Tests
Power up the device to get the oximeter ready for measurement.
2. Power-On Self-Test
Before using the PULSE OXIMETER, you must verify that the PULSE OXIMETER is
working properly and is safe to use. Proper working conditions are verified each time
when the PULSE OXIMETER is turned on as described in the following
procedure. The verification procedure (POST) takes 2 to 3 seconds to complete.
Caution: If any indicator or display element does not light when the PULSE
OXIMETER is turned on, do not use the PULSE OXIMETER. Instead, contact
qualified service personnel, your local MED-LINKET representative, or MED-LINKET's
Technical Services Department.
Section 2- Operation
Figure 4: Installing Batteries
Battery rear cover
English 11
Qty
1 Piece
1 Piece
1 Piece
1 Piece