Section 6- Service and Maintenance

Cleaning and Disinfecting

1. Clean the surface of the oxmeter by using a soft cloth dampened with either
a commercial, non-abrasive cleaner or a solution of 70% isopropyl alcohol in water,
and wiping it lightly the surfaces of the oximeter.
Please switch off pulse oximeter before cleaning. Clean the LED and
photo-sensor with moist cloth or cotton ball and alcohol gently.
2. The aforementioned general cleaning process is not for infection prevention.
Please contact the specialist for process of contagious infection.

Calibrating

1) Please use the SpO2 simulator of Fluke Biomedical index 2 to calibrate PULSE
OXIMETER for the function of SpO2 measurement. The calibration must be
operated to by qualified personnel only.
2) Please use the Temperature simulator of BC Biomedical MULTI-PARAMETER
PATIENT SIMULATOR to calibrate PULSE OXIMETER for the function of temperature
measurement. This calibration must be operated by qualified personnel only.
3) The SpO2 accuracy can be validated in human studies against arterial blood
sample reference measured with a CO-oximeter. All of the process of the clinical
study must be complied with standard of EN ISO80601-2-61.

Repairing and Maintenance

1. Please change the batteries when the low-voltage indicator lightens.
2. Please clean the surface of the device before using. Wipe the device with
alcohol first, and then let it dry in air or clean it by dry clean fabric.
3. Please take out the batteries if the oximeter is not in use for a long time.
4. The best storage environment of the device is -10℃ to 40℃ ambient
temperature and not higher than 80% relative humidity.
5. Please maintain properly for ensuring the device can be used normally.
6. The device needs to be calibrated once a year (or according to the calibrating
program of hospital). It can also be performed at state-appointed agent or just
contact us for calibration.
Warnings
● High-pressure sterilization cannot be used on the device.
● Do not immerse the device in liquid.
● It is recommended that the device should be kept in a dry environment.
Humidity may reduce the using life , or even damage the device.

Disposal

1) Used batteries should not be disposed of in the household rubbish. Used
Batteries should be deposited at a collection point.
2) At the end of its life, the appliance should not be disposed of in household
rubbish. Enquire about the options for regulations into account.