Transport and Storage
Products must be handled with care. Do not throw or drop them.
•
Protect the angled screwdriver and the accessories against mechanical damage.
•
•
Store and transport the products in secure containers / packaging.
Store the products under clean, dust-free and dry conditions.
•
The shelf life of the sterile goods depends on the storage conditions and type of packaging.
•
We recommend the wire basket (REF 50-990-40-04) and storage for accessories
(REF 50-990-41-04):
When returning the products please send back only cleaned and disinfected products in
•
sterile packaging.
8
Restriction Regarding Reprocessing/Maintenance
Frequent reprocessing has only minor influence on the service life of the angled screwdriver
and the accessories, which is mainly a function of use-related wear and tear or damage.
We recommend you carry out a regular service after 1000 sterilization cycles or 1 year.
Maintenance by the user is limited to the check of damaged parts, the oil service and the test
run.