The Spinal Modulation Neurostimulation System is intended for use in the electromagnetic
environment specified below. The customer or the user of the Spinal Modulation Neurostimulation
System should assure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
Recommended separation distances between portable and mobile RF communications
equipment and the Spinal Modulation Neurostimulation System
The Spinal Modulation Neurostimulation System is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the Spinal Modulation
Neurostimulation System can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the System.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
GUIDANCE AND MANUFACTURER'S DECLARATION
Electromagnetic Immunity
Compliance
Level
Level
3 V
3 V/m
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
0.12m
0.37m
1.17m
3.70m
11.70m
Electromagnetic Environment Guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of Spinal Modulation Neurostimulation System,
than 0.2 meter, based on transmitters of 80 MHz
to 2.5 GHz.
Interference may occur in the vicinity of
equipment marked with the following symbol:
80 MHz to 800 MHz
0.12m
0.37m
1.17m
3.70m
11.70m
19
800 MHz to 2.5 GHz
0.23m
0.74m
2.33m
7.37m
23.30m