EN – ENGLISH
It is possible to convert some of these older model clamps to accept the longer post of the
Modified Swivel Adaptor. Evaluations on suitability for conversion are made by the Service
Department.
Phone: 877-444-1114 (USA only) or (513) 533-7979
Integra LifeSciences Corporation
4900 Charlemar Drive, Building A
Cincinnati, Ohio 45227
Cleaning and Sterilization
The Modified Swivel Adaptor should be thoroughly cleaned after each use. Scrub with a
soft brush and use pH-neutral detergent. Clean thoroughly to prevent traces of blood or
debris from interfering with movement. Wipe down with an antiseptic solution
DO NOT STEAM STERILIZE!
Plastic components may be damaged by the heat.
Manual Wash
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause damage to the devices.
• Channels and crevices found on this device require particular attention during cleaning.
• Pay special attention to the water quality used throughout reprocessing. Hard water can
damage the surface of the equipment. Avoid using hard water. Instead use purified water
unless otherwise specified.
Limitations on reprocessing
• Repeated processing has minimal effects on these devices. Product life is normally
determined by wear and damage due to use.
• It is important to have Integra NeuroSpecialists perform routine inspections (twice yearly
is recommended). See contact information below.
INSTRUCTIONS
Containment/Transportation
• Follow health care facility protocol for safe containment and transport to the
decontamination environment.
• It is recommended that devices are cleaned immediately after use.
A close inspection of
the hole will reveal the
obstruction that
prevents the Modified
Swivel Adaptor from
operating with this
clamp.
10