4. Additionally the probe shall be suspended with the tip located no lower than the line
across the chamber (see Figures 10 & 11)
5. If a cable management system has been installed on your device, (see section A4.2),
secure the external portion of the cable and connector by:
Placing the connector carefully inside the tray.
Coiling the cable neatly and safely around the cable holder on the side of the tray
assembly.
NOTE: Incorrect positioning of the probe may result in:
High-level disinfection will not be achieved during the trophon EPR disinfection cycle
Residual disinfectant remaining on the probe surface. This may lead to temporary
bleaching and/ or irritation of the skin if no gloves are worn.
Damage to the probe may occur if in contact with the chamber wall
NOTE: Curved Probes must be correctly inserted in the device. Refer to Figure 12 for incorrect
and correct positioning of probe.
®
trophon
EPR User Manual – USA/Canada
Figure 12
L00035 1.0 03/2014
17