PART E – SERVICE AND WARRANTY
If you have any questions or concerns regarding the trophon EPR, trophon Printer or
disinfectant supplies, contact your GE Healthcare representative.
The trophon EPR and trophon Printer has a comprehensive warranty against defects in
material and workmanship for 12 months from the date of delivery.
Please ask your GE Healthcare representative for details about your trophon Printer warranty.
A warranty registration form is enclosed within this manual. Complete the warranty
registration form including the date at the time of initial set-up and return to your customer
service representative. If you have not received a warranty registration form or would like to
obtain further information about post warranty contracts please contact your GE Healthcare
representative immediately.
Service Schedule: To ensure correct operation, the trophon EPR must be serviced by
authorized service personnel. Service Schedule is every 12 months or at every 5000 cycles;
this will be indicated by the device by displaying the ccreen message: DEVICE SERVICE DUE
prior to commencing a disinfection cycle. This message will be displayed weekly on the
screen until a service is performed. Service intervals can also be accessed via the system
information, located in the trophon EPR LCD menu. When required, a disinfection history log
can be downloaded from your trophon EPR device and printed via trophon Connect
software or provided by your service centre. Contact your GE Healthcare representative to
obtain a print out for your records.
NOTE: Unauthorized modification to the trophon EPR or service by someone other than
authorized service personnel will void your warranty.
APPENDIX 1: TROPHON EPR TECHNICAL SPECIFICATIONS
Electrical Specification
Environmental
Specification
Physical Characteristics
EMC Compliance
®
trophon
EPR User Manual – USA/Canada
 Rated input voltage: 120V AC
 Rated input current: 5Amp, 50/60Hz
 Input IEC
 Output DB9
 Equipment must be connected to an earthed
outlet using power cable supplied with the
device
 Operating temperature range: 63
27
C)
0
 IP20
 Weight of device: 38 lb (17kg)
 Dimensions of device: 19.3in high x 13.6in wide x
13.6in depth (490mm high x 345mm wide x
345mm depth)
This device has been tested and found to comply
with the limits for emission requirements (Electro-
Magnetic Interference) pursuant to EN61326-
1:2006. CISPR 11 Group 1 Class A equipment to
CISPR 16-1 & CISPR 16-2, EN61000-4–2: 2005 & EN
61000-4-3: 2006.
L00035 1.0 03/2014