Intended Use:
The sole purpose of the trophon EPR is to high level disinfect validated ultrasound
probes (see section B2.1) according to the specified processes outlined in this manual.
It is not intended for any other use. Do NOT use this device for any application other
than its expressed purpose.
The trophon EPR together with the Sonex-HL is a high-level instrument grade
disinfectant system.
The trophon EPR is NOT intended to reprocess single use devices
The trophon EPR is NOT intended to pre-clean ultrasound probes
The cable management system is an accessory designed for use with the trophon EPR
(see section A4.2)
Chemical Indicator use is required with every disinfection cycle. More information can
be found in the Chemical Indicator Instructions for use, provided with the Chemical
Indicator.
The trophon Printer, trophon Connect, trophon Wall Mount and the trophon Cart are
accessories designed for use with the trophon EPR; contact your customer service
representative or visit the Nanosonics website for additional information. Refer to Part D
for information on the trophon Printer.
Disinfection Process:
At the beginning of the cycle, the trophon EPR creates an aerosol of concentrated
hydrogen peroxide. This is quickly and evenly distributed over the surface of the probe,
including very small crevices. This process provides thorough, high level disinfection of the
shaft and the handle of the probe. The device breaks down the hydrogen peroxide into
small amounts of water and oxygen and safely vents them into the external environment.
Notes:
1) Sonex-HL is the product name of the trophon Disinfectant.
2) The contact conditions are fixed cycle parameters that are not able to be modified by the
end user.
®
trophon
EPR User Manual – USA/Canada
L00035 1.0 03/2014
5