5 Performing the test
What to do
8. Aerate the sys-
tem
9. End the self-test Once the system has been
5 Performing the test
5.1 Test preparation
Before the test, you must decide whether a dynamic or static test is to be performed.
52
Operating manual
How to do it
Once the component test has
been completed, the display
shows a prompt to remove
the cap.
Remove the cap and press the
"Start/Execute" button.
The system is aerated; the dis-
play shows the current pres-
sure in the measurement
chamber.
aerated, the display automati-
cally shows the result of the
test.
If an error occurs, the system
stops the procedure, an error
code is displayed (see Trou-
bleshooting, p. 72), and the
system is aerated.
To repeat the test, press the
"Start/Execute" button.
To go back to the Main Menu,
press the "Back" button.
Button to use
LED
green
LED
red
Display
Self Test
Please remove
blind plug
select
back
Self Test
-20.10 hPa
back
Self Test
PASS
restart
back
Self Test
Error: 1XXX
See instruction manual
restart
back
PLT unit 997960