5 Performing the test
e.
f.
Changing the 4-channel adapter
5.3 Testing single-channel pipettes
Example: 100 μL air-displacement pipette, dynamic test
Before performing the test, set the pipette to the nominal volume. Mount a new unused disposable
tip, and insert the pipette with the tip into the previously selected and attached Adapter 1. Select an
adapter; see Test preparation, p. 52.
56
Operating manual
e.
Insert the selected pipette adapter into the adapter
housing (Adapter 2 for pipettes without tips shown
here).
f.
Screw the lock nut in tightly.
g.
After changing the adapter, perform the Quick Test.
The 4-channel adapter consists of an adapter with an inte-
grated lock nut. The entire ensemble is exchanged.
PLT unit 997960