What to do
How to do it
1. Call up the Main
Press or turn the select knob
Menu
to call up the Main Menu.
2. Set up the self-
Turn the select knob to the
test
"Pipette Type" setting.
3. Confirm the
Press the "Execute" button.
pipette type
The "Channels" submenu is
called up.
4. Select the Quick
Turn the select knob to the
Test
"Single Channel" setting.
5. Confirm the sin-
Press the "Execute" button.
gle-channel
The "Nominal Volume" menu
pipette
is called up.
6. Select the
Turn the select knob to the
pipette volume
"100 μl" setting.
7. Confirm the
Press the "Start/Execute" but-
pipette volume
ton. The start submenu is
called up.
PLT unit 997960
Button to use
5 Performing the test
Display
Main Menu
Pipette Type
Self Test
Switch OFF
Make settings
select
back
Main Menu
Pipette Type
Self Test
Switch OFF
Make settings
select
back
Channels
Single Channel
Multichannel
select
back
Channels
Single Channel
Multichannel
select
back
Nominal Volume
25 µl
50 µl
100 µl
200 µl
select
back
Nominal Volume
25 µl
50 µl
100 µl
200 µl
select
back
ready
SC - 100µl
back
select
Operating manual
57