OBJ_DOKU-163748-011.fm Seite 30 Freitag, 4. September 2015 8:35 08
7.7
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Move the movable parts back and forth (at least 5 times).
3. Mount the irrigation tube on the shaft tube.
4. Brush the inside of the shaft tube with a round brush (at
least 5 times).
5. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
7.8
Manual cleaning and disinfection
Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 12.3 (handles: max. 11!). Prohibited ingredients: organic
solvents, oxidizing agents, halogens, aromatic/halogenated hy-
drocarbons.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
1. Prepare a cleaning bath and a separate disinfecting bath ac-
cording to the manufacturer's instructions.
30