4. Press the rear release button (1) on the handle and slide the
shaft tube as far as it goes and latch into the handle.
6.4
Connect up the product
1. Connect the product to the multifunctional receptacle of the
Erbe VIO.
These products possess an electronic identifier (i.e. the product
and its specifications are recognized by the unit after the con-
necting cable has been inserted, and several default settings are
executed automatically on the unit).
6.5
Conducting a performance test
1. Moisten a gauze strip with a sterile NaCl solution.
2. Grasp the gauze strip with the instrument jaws and activate
the HF current.
Steam will form after approx. 2 to 3 seconds if the instrument is
working properly.
If damaged, do not use this product!
6.6
Using the product
The instrument is ready for use with 5 mm trocars. To insert into
a trocar sleeve, the instrument jaws must be closed.
To close the jaws, press the handle together. To open the jaws,
pull the handle apart.