Specifications (REF: 66801280/66801281/66802134)
Vacuum
Continuous Therapy Levels
Intermittent Therapy Levels
Intermittent Therapy Cycle Times High: 3, 5, 8, 10 minutes
Alarms
General (all alarms)
Priority
Auditory Sound Level
Indicator Color
Overall alarm delays
Over vacuum
High vacuum
Continuous Mode
Intermittent Mode
Low vacuum
Leak
Blockage
Canister full
Low battery
Critical battery
Battery failed
Inactive
Annual maintenance
Device failed
Power Requirements
Device input voltage
Device input power
Mains Adapter
Physical
Dimensions
Weight
25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg
High: 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg
Low: 0, 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180mmHg
Low: 2, 3, 5, 8, 10 minutes
Low
Low:
60 dB,
Medium: 68 dB,
High:
74 dB (1 meter from device)
Yellow
less than 5 seconds
180 seconds
60 seconds
60 seconds
45 seconds
120 seconds +/- 5 seconds
45 seconds +/- 5 seconds
immediate
immediate
30 seconds
15 minutes
Immediately following startup sequence
30 seconds
19.5 VDC
50 W
Smith & Nephew REF 66801286
Input: 100-240VAC, 50/60HZ, 10-35VA
Output: 19.5VDC, 2.6A, 50W
Fuses: Internal electronic fuse, not user changeable
180 x 190 x 76 mm (7" W x 7.5" H x 3" D) with 300ml canister
1.1 kg (2.4 lbs) with 300ml canister
17