Purpose
This Service Manual provides guidance on regular
service and maintenance checks to be conducted by
trained personnel on the RENASYS™ TOUCH device
(REF 66801280/66801281/66802134).
Note: The service and maintenance checks intended
at a facility level do not require opening the device.
Opening or attempting to open case will void all
warranties and render the user responsible for
all subsequent liabilities associated with device.
Contact your local Smith & Nephew customer care
representative or sales representative to resolve device
problems.
Scope
This service manual is only applicable to the
RENASYS TOUCH device manufactured by
Smith & Nephew (REF 66801280/66801281/66802134)
Responsibilities
It is the responsibility of the facility that owns or
operates the device to perform regular service and
maintenance checks of medical equipment to verify
basic device safety and function. The checks specified
in this manual are to be performed on a device at a
minimum once a year. However, these checks must
also be performed prior to any reuse or after prolonged
storage of more than 6 months. The retention of any
device service and maintenance records are the
responsibility of the party performing the procedure. A
sample checklist is included with this manual to provide
guidance on documenting the recommended service
and maintenance checks; however, the service and
maintenance documentation should be managed at the
discretion of each facility.
Training
It is recommended that individuals performing service
and maintenance on the RENASYS TOUCH device
be technicians familiar with regular service and
maintenance of medical equipment and with the use of
test equipment required.
Equipment and materials required
The following equipment is recommended to support
service and maintenance of RENASYS TOUCH devices.
Availability and usage is at discretion of each facility.
Note: All measurement/test equipment must be
calibrated and be in acceptable condition prior to use.
• RENASYS TOUCH canister
(REF 66801273, 66801274, 66801275)
• Quick click adapter (REF 66801259, included with
drain kits: REF 66801251,66801252, 66801253,
66801254, 66801255, 66801256, 66801257,
66801258)
• RENASYS Service Pack US ONLY (REF 66021812)
• RENASYS Y-connector (REF 66800971, 66020971)
• RENASYS TOUCH Odor filter (REF 66801284)
• RENASYS TOUCH O-ring (REF 66801283)
• Vacuum measurement gauge with minimum range
of up to 250mmHg vacuum (e.g., Extech
406800 or 407910, Digitron
Series 475 or equivalent)
• Optional electronic data acquisition and printer for
checklist
• Minimum of protective gloves and eye protection
consult cleaning agent MSDS
• Vented hood, or local SOP guidelines for handling
bio hazardous materials
• Plastic O-ring removal tool (http://www.prepol.com/
product-range/product-subpages/o-ring-removal-
tool or http://www.golemgear.com/p-302-o-ring-
pick-plastic.aspx)
• #1 Pozidriv
screwdriver
®
• 5/32" (4mm) Slotted Screwdriver
• Stopwatch
• Cotton swabs (e.g., Q-tips
• Disposable paper towels or cloth
• Cleaning materials validated for use with
RENASYS TOUCH are:
· Neutral (pH 7) based detergent
· Diluted 10% (chlorine) bleach in water solution
· Germicidal/antibacterial treated disposable wipes
moistened with 70% Isopropyl Alcohol (70% IPA)
· Intermediate disinfectant agents (such as
Sporicidin
Disinfectant) that are safe for use
®
on plastic may be used. Follow manufacturer's
instruction carefully.