Operation check
1. Connect RENASYS™ TOUCH canister to device by
gently pushing canister over device inlet port and
engaging both canister clips.
2. Connect canister tubing quick click connector to
quick click adapter included with drain kits and
engage tethered cap into open connector of quick
click adapter.
3. Press Therapy Mode toggle icon on the homescreen
to select Continuous
4. Adjust therapy set point to 80mmHg by pressing
Increase
or Decrease
5. Press Start/Pause Therapy
6. Confirm device begins operating: status indicator
illuminates green, therapy Indicator
of touchscreen rotates orange and touchscreen
displays – "Leak Check", "Seal Achieved" and
"Delivering Therapy".
Engaged
tethered cap
therapy.
icons.
button.
at the top
9
7. Press and hold Lock/Unlock
8. Confirm user interface is locked by pressing Start/
Pause Therapy
button and confirming device
does not stop operating.
9. Press and hold Lock/Unlock
2 seconds.
10. Stop device operation by pressing Start/Pause
Therapy
button. The status indicator turns off,
therapy indicator turns gray and device returns to
Home screen.
Operation on Battery check
1. Confirm that the therapy set point is at 80mmHg
and disconnect power supply from device.
2. Press Start/Pause Therapy
3. Confirm device starts operating: status indicator
illuminates green and after leak check is complete
the "Delivering Therapy" screen displays.
4. Press Start/Pause Therapy
device stops operating: status indicator turns off,
therapy indicator turns gray and device returns to
Home screen.
button for 2 seconds.
button for
button.
button and confirm