Service and maintenance performed after
each patient use
Device cleaning
Recommended personnel: Device cleaning should be
performed by a trained technician.
Equipment and materials required: Gloves, eye
protection, neutral (pH7) based detergent, 10% chlorine
bleach solution, germicidal/antibacterial treated
disposable wipes, cotton swab and dry disposable
cloth.
Precautions:
• Use of gloves and eye protection or other personal
and environmental protections is recommended.
The party conducting any cleaning should consult
the cleaning agent MSDS and their facility or local
operating procedures for specific requirements for
handling biohazards.
• Device should not be connected to AC mains power
while being cleaned.
• Follow manufacturer's guidelines for safe use and
disposal of cleaning materials.
• Do not spray cleaning agent, immerse any part of
device in fluid or use an excessively wet cloth.
Cleaning instructions
1. Place device and all associated accessories under
a vented hood. This includes all items that may be
associated with use of the device: power supply,
power cord, IV pole/bed clamp, etc.
2. Remove device from plastic bag, if used, and
discard bag.
3. If canister is present, remove it from device and
place in a plastic bag for disposal in accordance
with facility protocols and/or local ordinances for
disposal of potentially infected or bio-hazardous
materials.
4. Place other single patient accessories, such as
the carry bag or carry strap, in a plastic bag and
dispose in accordance with facility protocols or local
ordinances.
5. Wipe down device surface with a dampened cloth
or disposable wipe. Carefully clean all external
surfaces of device with one of the recommended
cleaning materials
6. Check that inlet port and air exhaust outlet are free
of obstructions, dust or foreign materials.
7. Apply germicidal/antibacterial product to a cotton
swab and clean both inlet port and air exhaust outlet.
8. Visually inspect device surface for debris or soils
that have not been removed and repeat cleaning
steps if necessary.
9. If RENASYS™ TOUCH IV pole/bed clamp (66801278)
is present
• Wipe down clamp surface with a dampened
cloth or disposable wipe.
• Turn clamp knob to fully extend clamp.
• Open clamp clips.
• Carefully clean all external surfaces with one of
the recommended cleaning materials.
• Clean under clamp clips and between lead
screw threads.
• When dry, turn knob to fully close clamp.
10. Wipe down power supply and power cord surfaces
with a dampened cloth or disposable wipe.
Carefully clean all external surfaces with one of the
recommended cleaning materials.
11. If damp, dry all surfaces with a disposable towel or
cloth.
12. Dispose of cleaning materials as instructed by the
manufacturer.
6