Functionality and alarm checks
Recommended personnel: Functionality and alarm
checks should be performed by a trained technician.
Equipment and materials required:
RENASYS™ TOUCH canister, RENASYS TOUCH power
supply, RENASYS power cord, RENASYS quick click
adapter and stopwatch.
Functionality check
1. Connect power supply to device power jack.
Connect power cord to power supply and to AC
mains power.
2. Confirm external power indicator illuminates green.
3. Turn on device by pressing and holding Power
button for 2 seconds. Confirm:
• status indicator illuminates while the power
button is pressed
• Smith & Nephew start up screen displays
• device issues an audible tone, then
• Welcome screen displays
The Home screen will display upon completion of
startup sequence.
*smith&nephew
4. Confirm device powers on and is ready to accept
user inputs.
5. Confirm battery status indicator on touchscreen has
the battery is charging.
6. Press Settings
and scroll to Device Information.
Note: The touchscreen should only be actuated by finger.
Using pens or other pointed objects may damage the
screen. Tap touchscreen to make a selection. Slide finger
to scroll.
7. Select device information and locate software
versions.
8. Record Serial Number, Operating System, Main
Application Version, Motor Controller Version,
Bootloader Version, RENASYS TOUCH System
Version, Battery Remaining and Device Lifetime
Therapy displayed on device information screen.
9. Press Back
icon to return to the Settings menu.
10. Device language may be changed as described
below:
• In the Settings menu, scroll to Language.
• Press Language to select.
• Scroll and select preferred language from
available list.
• Press Back
icon to return to Settings menu.
Clinician mode
To perform operation checks, device must be in
Clinician mode.
• In the Settings menu, scroll to Change Mode.
• Press to select Change Mode.
• If Patient mode is highlighted, press Clinician mode
mode.
• If Clinician mode is highlighted, no change is
necessary.
• Press Back
icon to return to Settings menu.
• Press Home
icon to return to Home screen.