Clear the overcurrent condition
The OCP trip can only occur if the corresponding output is enabled. When the
OCP trip occurs, the corresponding output is disabled automatically.
– If the OCP trip occurs on Output 1, the display shows triP on line 1 and the
– If the OCP trip occurs on Output 2, the display shows triP on line 2 and the
1 Select the output channel with the OCP trip.
2 Clear the OCP trip.
If the OCP trip persists, decrease the current limit settings to clear the tripping.
NOTE
Keysight U8030A Series User's Guide
OCP1 annunciator blinks.
OCP2 annunciator blinks.
– Press [1] to select Output 1, or
– Press [2] to select Output 2.
– The corresponding annunciator (OUT1 or OUT2) turns on.
– Press the [Over Current] button again to clear the OCP trip.