Operation and Features
Extending the Voltage and Current Range
– Never exceed the floating voltage rating of any of the supplies.
CAUTION
– Never subject any of the power supplies to negative voltages.
LETHAL VOLTAGES AND CURRENTS
WARNING
– Keep away from live circuits. This instrument operates up to 60 VDC at its
– SHOCK HAZARD. Floating voltage must not exceed 240 VDC. The total
– The output terminal is designed for DC application. Ensure that transient
– Personal injury on contact may result if these terminals are touched when
– To avoid injuries, always disconnect power, discharge circuits, and
– Installation procedures must be performed by a qualified and trained
Two or more power supplies can be connected in series or parallel to extend the
range of voltage and current. This may serve as a lower cost alternative to a
power supply with higher power rating.
60
output terminals and sense terminal.
voltage on the output terminal and the floating voltage must not be more
than 240 VDC from chassis ground.
voltages do not exceed 480 VPK or 480 VDC.
equipment is in operation.
remove external voltage sources before any installation.
personnel on this equipment.
Keysight U8030A Series User's Guide