Descargar Imprimir esta página

Chauvin Arnoux SIMPLE LOGGER II Serie Manual De Usuario página 44

Registrador digital autónomo

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

SIMPLE LOGGER II
MODEL
Electrical
Input
Input connection
Input level
Uncertainties (50 / 60 Hz)
Resolution
Maximum input level
Input impedance
Sample rate
Storage rate
Storage modes
Recording length
Memory
Communication
Power source **
Battery life
Mechanical
Dimensions
Weight (with batteries)
Case
Vibration
Shock
Drop
Environmental
Operating temperature
Storage temperature
Relative humidity
Altitude
Safety
Conformity
Level of protection
*See the end of chapter
Two 4mm safety banana jacks
0 to 10 mA: unspecified
10 to 50 mA: ± (0,5% of reading + 1 mA)
50 to 1000 mA: ± (0,5% of reading + 0,5 mA)
64 samples/cycle
Programmable from 125 ms to 1 day
Start/Stop, FIFO or Extended Recording Mode* (XRM™)
Programmable using DataView
240,000 measurements (512 kB). The recorded data is stored in
non-volatile memory retained even if the battery is low or removed.
USB 2.0 optically isolated
2 x 1.5 V AA (LR6) cell Alkaline batteries
100 hours to >45 days (dependent on storage rate/recording length)
136x70x32 mm (5.38x2.75x1.28")
IEC 68-2-6 (1.5 mm, 10 to 55 Hz)
IEC 68-2-27 (30 G)
IEC 68-2-32 (1 m)
- 10°C to + 50°C (14°F to 122°F)
- 20°C to + 60°C (- 4°F to + 140°F)
up to 85% at 35°C (95°F) without condensation
EN 61010-1; 50 V CAT III; Pollution level 2
L111
One
0 to 1A
AC
0.1 mA
1,2 A
1 Ω
®
from 15 minutes to 8 weeks
180 g (6.4 oz)
UL94-V0
2000 m
IP 40
English
Page 44

Publicidad

loading