cially suited to the rated voltage of the BENNING CM 2. The measuring
tips can be protected by caps.
5.1.13 Widest prong opening: 25 mm
5.1.14 Largest wire diameter: 22 mm
6.

Ambient conditions

-
The BENNING CM 2 is designed only for measuring in dry surroundings.
-
Maximum barometric height during measurement: 2000 m
-
Overvoltage category / set-up category: IEC 60664-1/ IEC 61010-1 → 300 V
category III, 600 V category II,
-
Degree of contamination: 2
-
Protection Class: IP 30 (DIN VDE 0470-1 IEC/ EN 60529)
IP 30 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 2.5 mm, (3 - first index). No protec-
tion against water, (0 - second index).
-
Operating temperature and relative humidity for voltage and resistance
measurements:
At operating temperature of 0 °C to 30 °C: relative humidity under 80%
At operating temperature of 30 °C to 40 °C: relative humidity under 75%
At operating temperature of 40 °C to 50 °C: relative humidity under 45%
-
Operating temperature and relative humidity for current measurements:
At operating temperature of 0 °C to 30 °C: relative humidity under 80%
At operating temperature of 30 °C to 40 °C: relative humidity under 75%
-
The BENNING CM 2 can be stored at temperatures from - 20 °C to + 60 °C.
The batteries must be removed from the unit.
7.

Electrical data

Note: The measurement accuracy is stated as the sum of
-
a relative proportion of the measurement value and
-
a number of digits (i.e. numerical steps of the last place).
This measurement accuracy applies for a temperature of 23 °C ± 5 °C (23 °C ± 3 °C
for current measurement) and a maximum relative humidity of 80%.
The BENNING CM 2 has an automatic switch-over of measurement range.
Previous setting is therefore not required.
7.1 DC voltage range
The input resistance is ≥ 10 M.
Measuring range
400,0 mV
4,000 V
40,00 V
400,0 V
600 V
7.2 AC voltage range
The input resistance is ≥ 10 MΩ parallel 100 pF. The measuring value is gained
and indicated as effective value (True RMS, AC coupling). Its calibration is
adapted for a sinusoidal curve form. With deviations from this form, the display
value becomes less accurate. The additional error for the following crest factors
is as follows:
crest factor of 1.4 to 2.0: additional error ± 1.0%
crest factor of 2.0 to 2.5: additional error ± 2.5%
crest factor of 2.5 to 3.0: additional error ± 4.0%
Measuring range
400,0 mV
4,000 V
40,00 V
400,0 V
05/ 2015
Resolution
Measurement accuracy
100 µV
± (0,5 % of reading + 5 digits)
1 mV
± (0,5 % of reading + 2 digits)
10 mV
± (0,5 % of reading + 2 digits)
100 mV
± (0,5 % of reading + 2 digits)
1 V
± (0,5 % of reading + 2 digits)
Resolution
Measurement accuracy
± (2,0 % of reading + 5 digits)*
100 µV
in frequency range 50 Hz - 60 Hz
± (1,5 % of reading + 5 digits)
1 mV
in frequency range 40 Hz - 300 Hz
± (1,5 % of reading + 5 digits)
10 mV
in frequency range 40 Hz - 500 Hz
± (1,5 % of reading + 5 digits)
100 mV
in frequency range 40 Hz - 500 Hz
BENNING CM 2
Overload protection
600 V
eff
600 V DC voltage
600 V
eff
600 V DC voltage
600 V
eff
600 V DC voltage
600 V
eff
600 V DC voltage
600 V
eff
600 V DC voltage
Overload protection
600 V
eff
600 V DC voltage
600 V
eff
600 V DC voltage
600 V
eff
600 V DC voltage
600 V
eff
600 V DC voltage
12