Precaution: When drawer is extended from Console, avoid knocking or leaning on
the drawer. This may damage the device.
Load Pump
1.
Sterile technician: Remove the catheter and sufficient tubing for ease of use
from the sterile package and inspect for damage. Hand the tray with the rest
of the Thrombectomy Set to the nonsterile technician for installation into the
Console (Figure 5).
Figure 5�
2.
Nonsterile technician: Remove the rest of the Thrombectomy Set from the
tray by gripping the pump (not the piston head) and insert the pump into the
Console (Figure 6).
Figure 6�
3.
Ensure that the waste tubing aligns with the roller pump (Figure 7).
Saline
Delivery
Tubing
Figure 7�
4.
Remove the cap from the Thrombectomy Set bag spike and insert the spike
into the saline bag.
5.
Push the drawer button to close the Console drawer (Figure 8).
Figure 8�
The Console will load the pump and scan the bar code located on the pump. After the
Console successfully identifies the catheter, the status panel will display the model
of catheter in use (Figure 9).
7
Roller
Pump
Waste
Tubing
CMYK Image /
Black (K) ∆E ≤5.0