Aesculap
ELAN 4 electro micro sagittal saw GA836
7.8
Mechanical cleaning/disinfecting with manual pre-cleaning - tools
Note
The cleaning and disinfecting machine must be of tested and approved
effectiveness (e.g. FDA approval or CE mark according to DIN EN
ISO 15883).
Note
The cleaning and disinfection machine used for processing must be ser-
viced and checked at regular intervals.
Manual pre-cleaning with ultrasound and brush
Phase
Step
I
Immersion
II
Rinsing
III
Ultrasonic cleaning
IV
Cleaning with brush
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Helizyme
**Recommended: BBraun Stabimed
Phase I
►
Fully immerse the product in enzymatic cleaning solution for at least
30 minutes. Ensure that all accessible surfaces are moistened.
Phase II
►
Rinse the entire product (all accessible surfaces) under running water.
Phase III
►
Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for
at least 15 minutes. Ensure that all accessible surfaces are immersed
and acoustic shadows are avoided.
Phase IV
►
Clean the product with a suitable cleaning brush until all discernible
residues have been removed.
12
®
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold) >30
50
RT (cold) -
-
55/131
>15
2
RT (cold) -
-
Water
Chemicals
quality
D–W
Enzymatic detergent*
D–W
-
D–W
Concentrate-free, aldehyde-free, phenol-free, QUAT-free, pH ~ 9**
D–W
-