Descargar Imprimir esta página

HT SOLAR03 Manual De Instrucciones página 37

Ocultar thumbs Ver también para SOLAR03:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 44
22/03/23 – 15:35
Irr. F
Irr. BH Irr. BL Tmp/A
[Off]
[Off]
[Off]
22/03/23 – 15:35
HELP
INFO
STOP RECORDING
22/03/23 – 15:35
Stop recording?
(ENTER/ESC)
3. Activate the recording of SOLAR03 through the relevant control
available on the associated Master instrument (see the Master
instrument's user manual). The indication "REC" is shown on the display
as indicated in the screen to the side. Recording interval is always 1s
(cannot be changed). With this sampling interval it is possible to
carry out recordings with the duration indicated in section
"Memory"
REC
4. Bring the remote unit near the modules and connect the irradiance/
temperature probes. Since SOLAR03 will record all values with an
[Off]
interval of 1s, the Bluetooth connection with the MASTER unit is NO
longer strictly necessary
5. Once measurements carried out by the Master unit are complete, bring
the remote unit near again, wait for the automatic connection and stop
recording on the Master instrument (see the relevant user manual).
The indication "REC" disappears from the display of the remote unit.
Recording is automatically saved in the remote unit's memory (see §
5.2.2)
6. At any time it is possible to manually stop the recording of parameters
on the remote unit. Use the arrow keys  or , select control "STOP
RECORDING" as shown to the side and press key SAVE/ENTER. The
following screen appears on the display
7. Press key SAVE/ ENTER to confirm that recording should be
stopped. The message "WAIT" shortly appears on the display and
recording is automatically saved
In case recording is stopped from the remote unit, values of
irradiance/temperature will be missing for the measurements
subsequently carried out with the Master instrument, and therefore
measurements @STC will not be saved
CAUTION