6.6. DYNAMIC MEASUREMENT
PURPOSE
Operator can use this program to measure dynamic weight. The dynamic weighing way is summarize the weighing result from setting time and
average it
INSTANCE
Set 10 seconds for the dynamic weight material or variable weight material.
KEY (ORDER)
Press and Hold
[MENU] Key
Short Press [CAL] Key
Short Press [MENU]
Key Five Times
Short Press [CAL] Key
Short Press [CAL] Key
Short Press [CAL] Key
Press and Hold [TARE]
Key
Press and Hold [TARE]
Key
•
Quick restart: exit the present computing price and restart the new one. Press and hold (MENU) key can restart the step 1, short press
(CAL) to enter into step 3.
NOTE: The grey color words explain the signal's meaning which flash on the window.
STEP EXPLANATION
1. Enter Into Menu
2. Display counting mode
3. Enter Into Dynamic Measurement
The signal hash on the upper left of window
4. Select Weighing Time
•
Press [PRINT] key can cycle and select different weighing
time. (Second)
•
Operator can set the weighing time by: Press [UNIT] key to
move cursor, press [PRINT] to select the target Number.
5. Confirm the weighing time
•
Three position to display as:
Upper left display present weight, Upper right display the
setting time, Main window display: Start
6. When display flash: START, put weighing sample on pan
7. Start to weigh for 10 seconds
8. Average the weighing result automatically
after 10 seconds.
•
Three position to display as (Example: 98.423g):
Upper left display the dynamic value, Upper right display
the weighing time, Main window displays the average
value.
9. Clear the weighing data
•
(If need to measure different material, please repeat step
7-9.)
10. Exit the dynamic measurement
USER'S MANUAL ABS
MENU LEVEL
LCD SCREEN DISPLAY
AND CODE
--node-
-COUNT-
dYNAMIC
Td---
10
STArT
0.000 g
EN
1 .
1 .1 .
1 .6.
1 .6.1
85