Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux
appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions
suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de
l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est
susceptible d'en compromettre le onctionnement.
INTERNAL BATTERY INFORMATION
This Normatec Go control unit is equipped with a rechargeable lithium ion battery.
The internal battery is designed to allow use of the Normatec Go System anywhere—
even when power outlets aren't available. The Normatec Go control unit may need
to be charged before first use. A fully charged battery will provide power for up to
2 hours of cumulative run-time. It takes approximately 4 hours to fully charge the
battery when the control unit is plugged in and not in use. The rechargeable lithium
ion battery is intended to be changed only by authorized service personnel with the
use of a special service tool.
• This device contains a lithium-ion battery. Batteries are only replaceable by
a skilled worker with a special tool. Contact Hyperice for service or repair.
When the battery is at end of life, the device shall be properly disposed
of. The battery must be complied with safely at an appropriate e-waste
disposal or recycling facility.
PRODUCT SPECIFICATIONS
• Normatec Go Model: ALJ7
• Normatec Go Calf Wrap Dimensions (fully open): 25" (width), 1" (depth), 12"
(height); [63.5 cm (width), 2.5 cm (depth), 30.5 cm (height)]
• Normatec Go Weight: 1.2 lbs [0.5 kg]
• Normatec Go electrical requirement: 5V DC 1 A
• Maximum Air Pressure: 220 mm Hg
• Internal Battery Rating: 3.6V, 2550mAh/3.65V, 2600mAh
• Ambient Temperature (charging): + 50 °F to 95 °F [+10 °C to 35 °C]
• Temperature (operating): +41° F to 104° F [+5° C to +40° C]
• Temperature (storage): -13° F to +158° F [-25° C to +70° C]
• Relative Humidity (operating): 15% to 93%, non-condensing
• Relative Humidity (storage): -25˚ C without relative humidity control; +70˚ C
at relative humidity up to 93%, non-condensing
• Atmospheric pressure (storage and transportation): 190hPa to 1060hPa
• Atmospheric pressure (operating): 700hPa to 1060hPa
ELECTROMAGNETIC COMPATIBILITY
The information contained in this section (such as separation distances) is in general
specifically written with regard to the Normatec Go. The numbers provided will not
guarantee faultless operation but should provide reasonable assurance of such.
This information may not be applicable to other medical electrical equipment; older
equipment may be particularly susceptible to interference.
GENERAL NOTES
Medical electrical equipment requires special precautions regarding
electromagnetic compatibility (EMC) and needs to be installed and put into service
according to the EMC information provided in this document and the remainder of
the instructions for use of this control unit.