Immunity Test
Conducted
RF RF coupled
into lines (IEC
61000-4-6)
Radiated RF (IEC
61000-4-3)
EQUIPMENT CLASSIFICATION
• Protection against electric shock: Class II/internally powered equipment
• Degree of protection against electric shock: Type BF applied part
• Ingress protection: IP22
• Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide
• Continuous operation
TROUBLESHOOTING
Problem
The system does
not start
The wrap does
not inflate
The system
stopped pumping
Air leak message:
Air Leak
Compliance Level
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM bands
10 V/m
80 MHz to 2.7 GHz
Possible Causes
Power is not turned
on, battery is not
charged
The session has not
been started
The wrap is
punctured or
damaged
The session has
been stopped or
ended.
Air leak
Electromagnetic
Environment – Guidance
This device is suitable
for the electromagnetic
environment of typical clinic,
hospital, athletic training or
home environments.
Solutions
Press the power button to turn
the control unit on. Check that the
battery is charged.
Check that the start / stop LED
is ON and the timer is counting
down. If not, press start / stop
button to start or resume session.
Check that there is no air leak error
message on the display screen.
Check that the start / stop LED
is ON and the timer is counting
down. If not, press start / stop
button to start or resume session.
Check that there is no air leak error
message on the display screen.
Check that there is no Low Battery
message on the display screen.
Check for punctures in the wrap.