Descargar Imprimir esta página

Normatec Go Instrucciones De Uso página 12

Publicidad

Idiomas disponibles

Idiomas disponibles

Immunity
Against
Electrostatic
discharge, ESD
(IEC 61000-4-2)
Electrical fast
transients/
bursts (IEC
61000-4-4)
RF Proximity
(IEC 61000-
Surges on AC
mains lines (IEC
61000-4-5)
Power frequency
magnetic field
50/60 Hz (IEC
61000-4-8)
Voltage dips
and short
interruptions
on AC mains
input lines (IEC
61000-
4-11)
This equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of this equipment should assure that it is used in
such an environment.
Compliance Level
(of this control unit)
± 8 kV direct
± 2 kV ± 2 kV
Mains power quality
should be that of a
typical clinic, hospital,
athletic training, or
home environment.
± 2 kV
30 A/m
0.5 cycles 1 cycle 25
cycles (50
Hz) 30 cycles
(60 Hz) 250
cycles (50 Hz)
300 cycles
(60 Hz)
Electromagnetic Environment
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be kept at levels to reduce
electrostatic charge to suitable levels.
Mains power quality should be that
of a typical clinic, hospital, athletic
training, or home environment.
Equipment with high RF emissions
should be kept at a distance
to reduce the likelihood of
interference.
Mains power quality should be that
of a typical clinic, hospital, athletic
training, or home environment.
Equipment that emits high levels of
power line magnetic fields (in excess of
3A/m) should be kept at a distance to
reduce the likelihood of interference.
Mains power quality should be that
of a typical clinic, hospital, athletic
training, or home environment. If
you require continued operation
during power mains interruptions,
ensure that batteries are installed
and charged. Ensure that battery
life exceeds longest anticipated
power outages or provide additional
uninterruptible power source.

Publicidad

loading

Productos relacionados para Normatec Go