Chapter 6
|
Troubleshooting
6-2
|
Precision
Checklist
Bath will not start or shuts down
•
Check display for error messages.
•
Ensure
wasn't accidently pressed.
Ensure the circuit protector is in the on ( I ) position.
•
•
Check the line cord connection to your power supply and at the bath.
•
Make sure supply voltage is connected and matches the bath's nameplate rating ±10%.
•
Restart the bath.
No display
•
Pressing
should return the display.
•
Cycle the bath's circuit protector.
Inadequate temperature control
•
Verify the set point.
•
Low-end temperature set points require supplemental cooling.
•
Ensure bath installation complies with the site requirements in Chapter 3.
•
Make sure supply voltage matches bath's nameplate rating ±10%.
Please contact Thermo Fisher Scientific Sales Service and Customer Support if you need any
additional information, see inside cover for contact instructions.