The filters should be replaced every six months or after 500 cycles to maintain an adequate supply of
clean air during the air drying cycle.
To change the bacteria retentive air filter on the STATIM 2000 G4 and 5000 G4, follow these steps:
1. Power the STATIM OFF.
2. Disconnect tube A
remove the filter from the filter bracket
filter from the bracket, note the orientation of the arrow mark on
the filter.
3. When the filter is free of the bracket, carefully disconnect tube
B
4
from the filter.
4. Before installing the replacement bacteria retentive filter
(SciCan order no. 01-102119S) check that the arrow mark on the
filter matches the direction of the arrow on the bracket. Push the
left hand filter fitting into tube B
5. Gently press the replacement filter into the filter
bracket
. The arrow mark of the filter should be facing
3
out and pointing to the left.
6. Re-connect tube A
To change the air filter on the STATIM 2000 G4, follow
these steps:
1. Turn the power switch at the back of the unit OFF.
2. Remove and discard the old foam air filter
3. Install the new filter (SciCan part no. 01-100207S).
4. Secure the filter plate
disassembly procedure.
To change the air filter on the STATIM 5000 G4, follow these steps:
1. Unscrew the cylindrical air filter
2. Discard the old filter.
3. Screw the new filter (SciCan part no. 01-101652S), finger tight only, into place.
from the bacteria retentive filter
1
3
.
4
1
to the right hand filter fitting.
6
to the back of the compressor using the screw
8
in a counter-clockwise direction.
and
2
. As you remove the
2
5000 G4
.
7
Page 42
2000 G4
2
3
4
6
5
2
1
Figure 11
4
8
3
5
retained during the
1
7
Figure 10

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para SciCan STATIM 2000 G4

Este manual también es adecuado para:

Statim 5000 g4

Tabla de contenido