4
REPROCESSING IN A HOME ENVIRONMENT
The product components must be cleaned thoroughly immediately after each use, and disin-
fected once a week.
The connection tubing cannot be cleaned or disinfected.
Dry the connection tubing after each use [see: Care of the pressure gauge and connection
tubing, page 19].
The maximum operating life of the connection tubing is 1 year.
4.1 Reprocessing cycles
PARI PEP S System
Pressure gauge
4.2 Processing limits
PARI PEP S System, disinfection
4.3 Preparation
If a pressure gauge has been used:
• Detach the pressure gauge connector line from the connecting element.
• Detach the connection tubing from the pressure gauge.
• Detach the tubing adapter from the pressure gauge connector line and the connection
tubing.
General:
• If necessary, disconnect the PARI PEP S system from the nebuliser.
• Dismantle the PARI PEP S system into its individual parts.
• Open the cap on the connecting element.
16
– Clean immediately after every use
– Disinfect once a week
Cleaning and disinfection can be carried out together with
the nebuliser.
Care as required [see: Care of the pressure gauge and
connection tubing, page 19].
300 processing cycles, max. 1 year