Chemical disinfection
To ensure safety when handling chemicals, follow the instructions for use of the disinfectant.
EQUIPMENT:
The method has been validated in Europe using:
– Aldehyde-containing disinfectant: Bode Korsolex
Active agent basis: Aldehyde donor, aldehyde
– Application time: 30 minutes
PROCEDURE:
CAUTION
Risk of infection due to growth of bacteria
Inadequate disinfection encourages the growth of bacteria and thus increases the risk of
infection.
• Observe the mixing proportions indicated and the specified treatment time.
• Ensure that all components are completely submerged in the solution for the whole of
the treatment time. There must not be any air pockets or bubbles.
• Disinfect the individual parts with a solution prepared according to the manufacturer's in-
structions.
If the recommended application period is exceeded significantly, the plastic
parts may take on the smell of the medium used.
RINSING:
CAUTION
Risk of allergic reactions and irritation of the mucous membrane by disinfectants
Disinfectants can trigger allergic reactions or irritation of the mucous membrane on con-
tact with the skin.
• Rinse the product thoroughly to ensure that no residues of the disinfectant remain on
the PARI product.
• Rinse off all parts thoroughly in running water at about 15 °C for 3 minutes.
• Dispose of the used solution. The diluted solution can be disposed of down the drain.
DRYING:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean and absorbent surface, and allow them to dry completely.
26
®
basic (concentration: 4%)