Descargar Imprimir esta página

Pari PEP S System Instrucciones De Uso página 54

Ocultar thumbs Ver también para PEP S System:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 162
5.3 PARI PEP S System
Separated parts for processing
CAUTION
Risk of infection due to cross-contamination in the case of a change in patients
If a product is used for more than one patient, there is a risk that germs may be transmit-
ted from one patient to the next.
• Clean, disinfect and sterilise all separated parts before every patient change.
• Replace the connection tubing or carry out mechanical cleaning and disinfection of the
connection tubing [see: Connection tubing, page 28].
All components of the PARI PEP S System can be cleaned, disinfected and sterilised ac-
cording to the procedures described below.
The connection tubing must be treated separately.
Preparation
• Dismantle the product into its individual parts
Precleaning
All individual parts must be precleaned immediately after use.
EQUIPMENT:
– Drinking water temperature of about 15 °C
PROCEDURE:
• Rinse all parts used for 2 minutes in running drinking water.
Cleaning and disinfecting
Please observe the instructions for use for the chemicals used.
Manual cleaning
EQUIPMENT:
The method has been validated in Europe using:
– pH-neutral cleaning agent:
®
Bode Bomix
 plus (concentration: 0.1%)
– Application time: 10 minutes
22

Publicidad

loading

Este manual también es adecuado para:

018018g4200