5.4 Connection tubing
Mechanical cleaning and disinfecting
EQUIPMENT:
The method has been validated in Europe using:
– Alkaline cleaning agent: Dr. Weigert neodisher® MediClean forte
– Neutralising agent: Dr. Weigert neodisher Z
– Cleaning and disinfection device: RDG G7836 CD (Miele) (conforming to
DIN EN ISO 15883)
– Special baskets for Miele instrument dishwasher
– Compressed air source for drying
PROCEDURE:
Vario TD program or comparable valid programs
DRYING:
Dry the connection tubing as described in the section on this topic.
Drying
• Connect the connection tubing to a compressed air source (compressor or central medical
gas supply system).
• Switch the compressed air source on.
• Leave the compressed air source running until all the moisture in the tube has been re-
moved.
5.5 Visual inspection and storage
Check all individual components. Replace any broken, deformed or seriously discoloured
parts.
Storage location:
– dry
– dust-free
– protected from sources of contamination
optional: Use sterile packaging