Cleaning with disinfection
To ensure safety when handling chemicals, follow the instructions for use of the disinfectant.
Mechanical cleaning
with disinfection:
24
EQUIPMENT:
The method has been validated in Europe using:
– Alkaline cleaning agent:
Dr. Weigert neodisher
– Deionised water
– Cleaning and disinfection device as per DIN EN ISO 15883-1 and
15883-2.
Info: If a different alkaline cleaning agent is used, it may also be
necessary to use a neutralising agent. Follow the recommendations
of the manufacturer of the chemical.
PROCEDURE:
Program (min. A0 = 3000) for cleaning and disinfecting according to
manufacturer's instructions.
DRYING:
Ensure that there is no residual moisture remaining in the compon-
ents.
Shake the water out of all of the parts.
Place all parts on a dry, clean and absorbent surface and allow
them to dry completely.
®
MediClean forte (concentration: 0.5%)