Descargar Imprimir esta página

Beckman Coulter MicroScan RENOK Guia página 7

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 17
3.
Remove the rubber seal from the RENOK base and
discard in a biohazard waste container. Install a new
rubber seal and reassemble the RENOK in reverse order
of the preceding instructions.
Cleaning
Clean the RENOK only when contamination occurs. If it is
contaminated, the RENOK can be autoclaved.
Before autoclaving, apply a small amount of stopcock
grease to the rubber seal at the base of the unit.
Use non-fluid autoclave settings.
Do not clean by immersion.
Spare parts list
Rubber Seal (2)
Screw Kit (1 set)
Contact your Beckman Coulter representative for spare part
ordering information.
9020-7657, Rev. BD
Specifications
Nominal dispense volume per well:
Well-to-well repeatability, single panel:
Total dispense volume:
RENOK Accessories (inoculator sets)
D-Inoculator Sets (for Dried Conventional panels) and
R-Inoculator Sets (for Rapid panels) are IVD accessories and
include disposable seed troughs, and trough lids (vacuum
covers with pipette plates). See
component descriptions.
Intended purpose
All inoculator sets are intended for use only with the
MicroScan RENOK by laboratory professionals and are
single-use materials. Only use once per inoculum.
Summary and principles
MicroScan RENOK Disposable D-Inoculator Sets
(for Dried Conventional panels) are an in-vitro diagnostic
accessory which aids the RENOK in rehydrating and
inoculating MicroScan Panels for quantitative and
qualitative AST and ID bacterial identification.
Refer to MicroScan panel instructions for methods of
incubation and results determination which may include
manual, semi-automated, or full automation on
MicroScan instrumentation.
MicroScan RENOK Disposable R-Inoculator Sets
(for Rapid panels) are an in-vitro diagnostic accessory
which aids the RENOK in rehydrating and inoculating
MicroScan Panels for qualitative bacterial identification.
Refer to MicroScan Rapid Neg ID and Rapid Pos ID panel
instructions for methods of incubation and results
determination which may include manual, semi-
automated, or full automation on MicroScan
instrumentation.
See
Inoculation procedure (page 4)
using inoculator sets with the RENOK.
Inoculator set details
D-Inoculator Sets (for Dried Conventional panels)
Inoculator set, dried conventional, 240 count (B1013-4)
Inoculator set, dried conventional, 240 count, US (B1013-40)
Inoculator set, dried conventional, 60 count, US (B1013-5)
R-Inoculator Sets (for Rapid panels
Inoculator set, rapid, 240 count (B1013-6)
RENOK Rehydrator/Inoculator System Instrument Guide- EN
RENOK Rehydrator/Inoculator System
115 μL
± 10 μL
11.04 mL ± 0.96 mL
Introduction (page 3)