7.12 RESET FACTORY SETTINGS*
After enter setting mode, display shows "F0 CAL", press
key to reset factory settings.
8. BATTERY OPERATION
When battery voltage low, the battery indication (
soon as the arrow is on. To charge the battery simply plug into the AC adapter. The scale does not need
to be turned on.
There is an LED above display window to indicate the status of battery charging. If the LED is green the
battery has a full charge. If it is red indicates the battery is being charged.
As the battery is used it may fail to hold a full charge. If the battery life becomes unacceptable then con-
tact your distributor.
9. ERROR CODE
Error Message
Description
-----
Maximum load exceeded
Incorrect date
Err 1
Incorrect time
Err 2
Zero setting error
Err 4
Key board error
Err 5
A/D value out of range
Err 6
Percentage Error
Err 7
Calibration weight error
Err 8
Unstable Reading
Err 9
Wireless communiation failure
Err 10
Communication protocol error
Err 11
Accumulation error
Err 12
Lack of unit weight
Err 13
Lack of sample
Err 14
Gravity error
Err 15
Paper error
Err 16
Tare out of range
Err 17
Pre-tare error
Err 18
Initialize zero error
Err 19
Printer error
Err P
Approval setting error
Err l
Over range
Err--ol--
Underload
Err--lo--
key until display shows "F10 rt", press
) will turn on. The battery should be charged as
Solution
Unload or reduce weight
Enter the date by using format "yy;mm:dd"
Enter the time by using format "hh:mm:ss"
Zero setting range exceeded due to switching on.(4%max) Make sure
platform empty.
Check the keys and connecter.
Make sure platform empty and check the pan is installed proper.
Check the load cell connectors.
Please check input data, must be > 0.5d.
Check the test weights for calibration or linear calibration
Check any air variation, vibration, RF noise and touching some whe-
re. Check the load cell and connecters.
Check wireless settings or change Com settings from the wireless.
Check communication settings.
Max accumulation times 99 / 999 / weight 999999.
Check unit weight entry data, must be >0.5d.
Check counting samples entry, must be >20d.
Check the gravity settings. Gravity range must be 0.9xx~1.0xx
Check the printer paper.
Minus weight or overload. Remove the load and restart scale again.
Check the pre-tare value.
Calibration the scale.
Check the printer settings or connections
Check the PCB jumper settings. Must be connect
jumper pin to K1 (BW series)
Remove the load. Re calibrate
Minus weight, check the platform and restart or calibrate.