en
Troubleshooting
4.3
During use
1. Open the cylinder valve. The pressure gauge shows the
cylinder pressure of the device cylinder.
2. Put on connected masks.
3. The residual pressure warning whistle sounds from a
residual pressure of approx. 50 bar in the connected
device cylinder.
4.4
After use
1. Take off the mask.
2. Only close the cylinder valve when no more breathing air is
required.
3. Vent the lines via the lung demand valve.
4. Dismantle the pressure reducer unit from the device
cylinder.
5. Replace the used device cylinder with a completely filled
device cylinder.
6. Refit the pressure reducer unit onto the new device
cylinder.
7. Disconnect the PPE and replace it with cleaned and
disinfected PPE.
5
Troubleshooting
Error
Cause
Breathing air
Device cylinder is
extraction not
sealed.
possible
Connection to
device cylinder not
correct
The hose is kinked Remove the kink in
Cylinder valve
faulty
Gas leaking from
Connections are
the connection
not gas-tight.
points.
Leak at the relief
Malfunction with
valve
the pressure
reducer
Weak whistle
Residual pressure
sound
warning whistle
dirty
Residual pressure
Faulty starting
warning whistle
mechanism
does not work
correctly.
Couplings cannot
Couplings dirty or
be engaged or are
faulty
sluggish.
Breathing
Medium pressure
resistance is too
from the pressure
high.
reducer too low
14
Error
Usage period is too
short.
6
NOTE
Observe the maintenance intervals of the components, such
as the pressure reducer unit and the PPE used.
6.1
6.1.1
6.1.2
Remedy
Slowly open the
cylinder valve.
Ensure correct
connection to
6.1.3
breathing air
supply.
6.2
the hose and check
the hose for
6.2.1
damage. If
necessary, replace
1. Remove the PPE from the extraction points.
the hose.
2. Connect the filled device cylinder to the pressure reducer
Contact Dräger
Service.
3. Connect the quick connect coupling from the pressure
Contact Dräger
Service.
4. Open the cylinder valve so that the cylinder pressure is
5. Close the cylinder valve again.
Contact Dräger
6. After 10 minutes, no pressure drop must be visible on the
Service.
7. There must be no audible or visible leaks. If necessary, use
Clean and test
again.
Contact Dräger
8. After testing, prepare the product again according to the
Service.
6.2.2
1. Connect the filled device cylinder to the pressure reducer
Contact Dräger
Service.
2. Connect the quick connect coupling from the pressure
Contact Dräger
3. Open the cylinder valve so that the cylinder pressure is
Service.
4. Close the cylinder valve.
Cause
Device cylinder is
not completely
filled.
Leak
Maintenance
Maintenance intervals
Before use
Visually inspect the product for signs of external damage
and dirt
Every year
Functional test of the quick disconnect couplings
Leak test of the quick connect couplings and the manifold
block
Medium-pressure test of the pressure reducer unit
Check the sinter filter in the pressure reducer connection
for dirt or damage and replace if necessary.
Check the O-ring of the pressure reducer connection for
dirt or damage and replace if necessary.
Every 6 years
Basic overhaul of the pressure reducer at the manufacturer
Maintenance work
Leak test of the quick connect couplings and the
manifold block
unit.
reducer hose to the manifold block.
displayed on the inlet pressure gauge.
inlet pressure gauge.
leak detection spray for soaping. If the pipeline system
leaks, do not use the product until the system has been
repaired.
chapter "Before use" (see chapter 4.2).
Checking the residual pressure warning whistle
unit (cylinder pressure >70 bar).
reducer hose to the manifold block.
displayed on the inlet pressure gauge (>70 bar).
Remedy
Fill the device
cylinder completely.
Contact Dräger
Service.