1 6
SYSTEM
SYSTEM SPECIFICATIONS AND DIMENSIONS
Intelect® Legend 2 Head Unit
2 Channel COMBO
4 Channel COMBO
Cart configurations
Cart (Safe working load 6.5kg)
Device on cart
POWER
Input
100 - 240 V AC, 1.0 to 0.42 A, 50/60 Hz
Electrical Class
Mode of Operation
Note: Mains isolation is achieved by use of the double
pole switch located on the rear panel.
Electrical Type (Degree of Protection)
Ultrasound
Electrotherapy
ELECTRO STIMULATION SPECIFICATIONS
Output specifications are described for each waveform
from pages 18-21.
Unless otherwise specified, electrotherapy controls
accuracy is: ± 20 %.
Load impedance: 500-1000 Ohm
CC = constant current, effect of load impedance on
voltage
CV = constant voltage, effect of load impedance on
current
ULTRASOUND SPECIFICATIONS
Frequency
1 MHz +/- 10%; 3 MHz +/- 10%
Duty Cycles
Pulse Repetition Rate
Pulse duration:
INTELECT
LEGEND 2 USER MANUAL
®
Width
34cm
34cm
48cm (MAX)
-
CLASS I
Continuous
.TYPE B
.TYPE BF
10%, 20%, 50%, Continuous
100 Hz
1 -5 ms
Max (ON): 5 ms
Min (OFF): 5ms
Depth
Height
35.5cm
15cm
35.5cm
19cm
52cm (MAX)
96cm
-
111 cm
OUTPUT POWER
1cm
2
1MHz
3MHz
1
0.9
2W
1.8W
3W
2.7W
(*)
2W/
2W/
cm
cm
2
2
3W/
3W/
cm
cm
2
2
(*) An error of + 0.25 W can be measured with 1cm2 US applicator, pulse mode
100Hz at 10% or 20% Duty Cycle.
Unless otherwise specified, ultrasound
controls accuracy is:
Peak to Average Ratio:
Beam Nonuniformity Ratio
Beam Type
Treatment Time
Weight (no battery)
3.1kg
3.7kg
10.1kg
-
2cm
5cm
2
2
1MHz
3MHz
1MHz
3MHz
1.5
1
2.5
2.7
3W
2W
5W
5.4W
4.5W
3W
7.5W
8.1W
2W/
2W/
2W/
2W/
cm
cm
cm
cm
2
2
2
2
3W/
3W/
3W/
3W/
cm
cm
cm
cm
2
2
2
2
1:1, at 50% Duty Cycle
4:1, at 20% Duty Cycle
9:1, at 10% Duty Cycle
EN
10cm
2
1MHz
3MHz
6
6.8
12W
6.8W
18W
13.6W
2W/
1W/
cm
cm
2
2
3W/
2W/
cm
cm
2
2
± 20 %.
<5:1